Technical Parameter |
|
Technical parameters for impulse current immunity testing in accordance with IEC 61643-11 |
|
Single phase fully automatic coupling/decoupling network |
Compliant with EN/IEC 61643-11 GB/T
18802.11 |
Output current |
Up to 6 kA (8/20 us) |
Coupling network ports |
L1、N、PE |
Coupling path |
L1、N、 Any combination path of the route |
Coupling switching method |
Automatic switching |
Coupling device |
Direct coupling |
EUT carrying capacity |
Max AC 220 V(L1、N、PE)、16 A |
Max DC 220 V 16 A |
|
General parameters |
|
Scope of working power supply |
AC 110 V/220 V ± 10%, 50/60 Hz ± 5% (default AC 220 V 50 Hz in China) |
Fuse 200 W |
6A |
Maximum power consumption |
200 W |
External control mode |
25 needle parallel thread |
Instrument grounding connection method |
Use a flat grounding wire |
Cabinet size |
6 U |
Weight |
About 26.5 Kg |
Temperature range |
15-35 ℃ |
Humidity range |
45% - 75% |
Pressure range |
86 kPa – 106 kPa |
Standard accessories: Power cord, test wire, grounding wire, 25 core data cable, factory inspection report, instruction manual.
Optional accessories |
|
Impulse current simulator S6C600 |
Output 8/20 μ s pulse current waveform, with a maximum output current of 6 kA |