Introduction:
RSG 40C20 is a simulator for differential mode conducted disturbance immunity testing that complies with the requirements of IEC 61000-4-19/GB/T 17626.19 and JJF 1245.4-2019 standards. Supporting our company's Corelab software and various coupling decoupling networks, coupled transformers and other facilities can conduct the following tests: 2 kHz~150 kHz differential mode voltage and current (unmodulated waveform, continuous wave pulse with intermittent period, and pulse waveform modulated with rectangular pulse) tests that meet IEC 61000-4-19, and current tests that meet JJF 1245.4-2019.
Features:
> 5.7-inch touch screen;
> Compliant with IEC 61000-4-19/GB/T 17626.19 and JJF 1245.4-2019 standards;
> The integrated design makes the system more simple, safe and durable;
> Ethernet, RJ45 interface for PC remote control and test report print;
> The key components are all made of high-quality imported materials, with reliable quality, exquisite appearance and comfortable operation.
Technical Parameters |
|
Standard |
IEC 61000-4-19、JJF 1245.4-2019 |
Frequency |
2 kHz ~ 150 kHz |
Modulation frequency |
0~1000 HZ |
Frequency step |
Linear, logarithmic, static |
Technical parameters for differential voltage interference testing |
|
Wave form |
Sine, total harmonic distortion<5% |
Output voltage |
0.1 V ~ 40 V |
Load impedance |
10 Ω±30% |
Technical parameters for differential current interference testing |
|
Maximum current |
5A |
Load impedance |
1 Ω±0.3 Ω |
BNC output |
10:1(0.1 V / A) |
General Parameters |
|
Weight |
Approximately 25 kg |
Dimension |
620*450*190mm (length x width x height) |
Ambient Temperature |
15 ° C -35 ° C |
Relative Humidity |
45% -75% |
Atmospheric Pressure |
86 kPa ~ 106 kPa |
Standard Accessories |
|
Test line, power cord, fuse, factory report, instruction manual. |
|
Optional Accessories |
|
Coupling decoupling network CDN419M3-32 |
EUT voltage range: single-phase 3-wire, 32 A, 250 V AC/DC |
Frequency range: 2 kHz~150 kHz |
|
Coupling decoupling network CDN419M4N-32 |
EUT voltage range: three-phase 4-wire, 32 A, line to ground 250 V AC/DC |
Frequency range: 2 kHz~150 kHz |
|
Pulse transformer TPT-600-2 |
Coupling pulse voltage: Max 600 V |
Secondary side flow capacity: Max 30 A |
|
Used for IEC 61000-4-19 |
|
CoreLab anti-interference testing software |
Supports Windows 7 and above, it is easy to use, with a beautiful and intuitive user interface. Various operational functions and standard testing libraries enable users to easily complete custom testing programs; |
Users can customize any waveform (outside of the standard library); |
|
It can automatically/manually identify the connected CoreLab testing equipment and perform automatic configuration; |
|
Template based reporting function can help users flexibly generate test reports |
|
Differential current testing module WCT 20 |
Frequency range: 2 kHz~150 kHz |
Maximum current: 5 A |
|
Maximum voltage: < 22V |
|
Impedance: 1 Ω± 0.3 Ω |
|
Differential voltage testing module R419-10 |
Differential voltage test matching module, resistance value: 10 Ω |
Differential current testing module RJJF-100 |
100Ω |