Introduction
CPN 38100T30High voltage and high power impulse current coupling/decoupling network, coupled with host S15C3000 for 8/20μs current waveform test, coupled pulse current up to 30 kA. EUT maximum load capacity AC 380V 100A Three-phase five-wire, DC 380V 100A, the equipment is simple and fast operation, high reliability.
Features
> EUT Maximum load capacity AC/DC 380 100A;
> The coupled pulse current is up to 30kA;
> Easy to operate, automatic coupling mode;
> Ethernet LAN communication, stable and reliable.
Technical Parameters |
|
Three-phase Fully Automatic Coupling/Decoupling Network |
Accord with EN/IEC 61643-11、GB/T 18802.11 |
Coupled pulse current wave test |
|
Current up to 30kA (8/20μs) |
|
Note: The actual output pulse voltage value of the coupling/decoupling network depends on the set value of the pulse generator |
|
Current Output |
1.5KA ~ 30KA(±10%) |
Note: 1.5KA depends on the breakdown voltage of the gas discharge tube |
|
Coupled Network Port |
L1(DC+)、L2、L3、N(DC-)、PE |
Coupling Path |
L1(DC+)、L2、L3、N(DC-)、PE Line Any combination path |
Coupled Switching Mode |
Automatic switching |
Coupled Apparatus |
Gas discharge tube coupling |
EUT Carrying Capacity |
Max AC 380V 100A |
Max DC 380V 100A |
|
Accessories |
|
Display Screen |
5.7-inch TFT touch screen |
Operating Power Supply Range |
AC220V,(±10%),50Hz/60Hz |
Fuse |
10 A |
Maximum System Power Consumption |
1500 W |
Communication Mode |
Ethernet LAN |
External Control Mode |
25-pin parallel mouth thread |
Instrument Grounding Connection |
Use flat ground wires |
Dimensions |
22U*2 |
Weight |
Approx 410kg |
Environment Temperature |
15℃ - 35℃ |
Relative Humidity |
86 kPa - 106 kPa |
Barometric Pressure |
45% - 75% |
Optional Selection |
Test wire, fuse, power cord, ground wire, factory report, manual. |
Standard Equipment |
The CPN 38100T30 should be used with the impulse current simulator S15C3000. |