Technical Parameters |
|
Test Current Range |
100 A - 5000 A (± 10%) |
Current Waveform |
Wavefront time: 10 μs± 20% Half peak time: 350 μs ± 20% The wavefront time is defined as 0% -90% Half peak time is defined as 0% -50% |
Polarity |
Positive |
Source Output Impedance |
2.5 Ω |
General Parameters |
|
Display Screen |
10 inches color screen |
Working power supply range |
AC 220V, ± 10%, 50/60 Hz ± 5% (Domestic default AC 220 V 50 Hz) |
Fuse |
6A (MCS2000-6U)/10A (S12A500) |
Maximum Power Consumption |
1500 W |
Pulse Triggering Method |
Automatic |
Grounded Port |
Using a flat grounding wire |
Chassis size |
6U L * W * H: 550 * 280 * 450 (MCS2000-6U) 35U L * W * H: 800 * 600 * 1850 mm (S12A500) |
Instrument weight |
Approximately 15 kg (MCS2000-6U) Approximately 300 kg (S12A500) |
Environmental scope |
15℃ -35℃ |
Pressure range |
86 kPa - 106 kPa |
Humidity range |
45% - 75% |
Standard Accessories |
Test wires, Fuses, Power Lines, Grounding Wires, Factory Reports and Manuals. |