General Parameters |
|
Display Screen |
5.7 TFT touch screen |
Working Power |
AC 110 V/220 V(±10%),50 Hz /60 Hz(Default in mainland China:220 V) |
Fuse |
6 A |
Max. Power |
200 W |
Communication Mode |
Ethernet LAN、RJ45 |
Safety Loop |
Stop working when the safety circuit is not closed |
Failure Detection |
The failure information will be displayed on LCD in front panel when failure occurs and interrupt the working of instrument. |
Earthing Mode |
Flat earth line |
Dimension |
PVSG 1500 : 8U PVCL 1500A : 5U PVCL 1500B : 5U |
Weight |
PVSG 1500 : 44 kg PVCL 1500A : 29 kg PVCL 1500B : 30 kg |
Ambient Temperature |
15 ℃ ~ 35 ℃ |
Relative Humidity |
45% ~ 75% |
Atmospheric Pressure |
86 kPa ~ 106 kPa |
Technical Parameters |
|
Standard |
IEC 61730-2: 2016 MST14 |
Test Voltage |
0.5 kV ~ 15.0 kV |
Calibration voltage |
3 kV ~ 15.0 kV ±3%; |
Rise Time |
1.2 μs±30% |
Duration |
50 μs±20% |
Polarity |
Positive, negative |
Waveform forming network |
Following Load capacitance |
Capacitance range |
20 nF to 170 nF, total of 11 gears, manually selected (short plug) 20 nF – 25 nF , 22.5 nF ±3% 25 nF – 30 nF , 27.5 nF ±3% 30 nF – 35 nF , 32.5 nF ±3% 35 nF – 40 nF , 37.5 nF ±3% 40 nF – 50 nF , 45 nF ±3% 50 nF – 60 nF , 55 nF ±3% 60 nF – 75 nF , 67.5 nF ±3% 75 nF – 95 nF , 85 nF ±3% 90 nF – 115 nF , 105 nF ±3% 115 nF – 140 nF , 127.5 nF ±3% 140 nF – 170 nF , 155 nF ±3% |
Repetition Rate |
8 s ~ 9999 s (The shortest time depends on test voltage) |
Counter |
1~ 9999 |
Trigger Mode |
Manual, Auto, External Trigger |
Measurement |
Monitor via coaxial port on front panel: 10 V @ 15 kV |
Peak Value Acquisition |
Display on touch screen, MAX : 15 kV |
Low-voltage Protection |
When the pulse voltage does not reach the set value, the test process stops |
Charging Display |
Count-down display |
Standerd Accessories |
User Manual, Test line, short-circuit plug, Earth line, 3-core power line, fuses (Spare parts),PVCL 1500A&PVCL 1500B , Test Box |
Optional Accessories |
PC control software:Corelab Windows 7, Windows 8 and Windows 10 are supported, including various operating functions and standard libraries. Users can easily control the test equipment remotely, complete the test work, and automatically configure it to help users flexibly generate test reports. |