General Parameters |
||
Display Screen |
5.7” TFT touch screen |
|
Working Power |
AC 110 V/220 V(±10%),50 Hz /60 Hz(Default in mainland China:220 V) |
|
Fuse |
6 A |
|
Max. Power |
200 W |
|
Memory Space |
Infinite(PC) |
|
Communication Mode |
Ethernet LAN、RJ45 |
|
Failure Detection |
The failure information will be displayed on LCD in front panel when failure occurs and interrupt the working of instrument. |
|
Working Status Indication |
LED indication and LCD display on front panel |
|
EUT Power Injection Line |
φ4 mm banana plug |
|
Detection Output Mode |
Standard 1m coaxial cable |
|
HV Output Mode |
4 mm2 banana plug(HV line) |
|
Dimension |
19”/ 6U |
|
Weight |
Approx. 30 kg |
|
Ambient Temperature |
15 ℃ ~ 35 ℃ |
|
Relative Humidity |
45% ~ 75% |
|
Atmospheric Pressure |
86 kPa ~ 106 kPa |
|
|
Technical Parameters |
|
|
Combination Wave |
|
|
Test Voltage |
0.3 kV ~ 6 kV (±10%) |
|
Telecom Wave |
|
|
Test Current |
7.5 A-150/200 A (±10%) |
|
Voltage Waveform |
Rise time :10 μs±30%, time to half value:700 μs±20% |
|
Current Waveform |
Rise time : 5 μs±20%, time to half value: 320 μs±20% |
|
Source Impedance |
15 ohm /40 ohm (Switched automatically) |
|
Polarity |
Positive, negative, alternative |
|
Pulse Repetition |
6 s - 180 s (The shortest time depends on test voltage) |
|
Counter |
1–999 |
|
Trigger Mode |
Manual, Auto, External Trigger |
|
CDN |
External CDN model No. (options): CDN-405AF8 CDN-405T8 CDN-405T8A1 |
|
Output voltage up to 8 kV and 10 kV are available for customization. |