General Parameters
|
Display Screen
|
5.7” TFT color touch screen
|
Power Supply
|
AC 110 V/220 V(士10%), 50
Hz /60 Hz,(士5%) (default AC 220 V 50 Hz in mainland China)
|
Fuse
|
6 A
|
Max. Power Consumption
|
200 W
|
Memory Space
|
Infinite(PC)
|
Communication Mode
|
Ethernet, RJ45
|
Dimension
|
19”/6U
|
Weight
|
Approx. 35 kg
|
Ambient Temperature
|
15 ℃-35 ℃
|
Relative Humidity
|
45%-75%
|
Atmospheric Pressure
|
86 kPa-106 kPa
|
|
|
EFT/burst Tests as per IEC 61000-4-4
|
Test Voltage
|
0.25 kV-4.8 kV(±10%)
|
Source Impedance
|
50 Ω
|
Polarity
|
Positive, negative, alternative
|
Rise Time
|
5 ns (±30%), 50 Ω load, 1000 Ω load
|
Pulse Waveform
|
5/50 ns, 50 Ω and 1000 Ωload
|
Pulse Frequency
|
0.1 kHZ-1000 kHz
|
Pulse Duration
|
50 ns(±30%),50 Ω load
|
Burst Duration
|
0.075 ms-750 ms
|
50 ns-15/+100 ns,1000 Ω load
|
Test Duration
|
1 s-9999 s
|
Burst Cycle
|
11 ms-9999 ms
|
Synchronization
|
0°-360°, 1° step or random
|
Trigger Mode
|
Automatic, manual or external trigger
|
CDN
|
Built-in single-phase automatic CDN:
Single-phase-3-line
AC 220 V 16 A Max. 250 V
|
Surge Tests as per IEC61000-4-5
|
Test Voltage
|
0.3 kV-5 kV (±10%)
|
Test Current
|
0.15 kA-2.5 kA (±10%)
|
Voltage Waveform
|
Front time: 1.2 μs (±30%), Time to half
value: 50 μs (±20%)
|
Source Output Impedance
|
2 Ω
|
Current Waveform
|
Front time: 8 μs (±20%), Time to half
value:20 μs (±20%)
|
Coupling Capacitance
|
9 μF, 18 μF
|
Polarity
|
Positive, negative, alternative
|
Pulse Cycle
|
5 s-180 s (the shortest time depends on
test voltage)
|
Coupling Resistance
|
10 Ω, 0 Ω coupling resistor Either IEC
standard mode or customized mode is optional
|
Number of Test
|
1–999
|
Surge Voltage and Current Peak
Detection
|
Front panel BNC output, surge voltage: 1000 V:1 V Surge current :500 A:1 V 3 digits on display
|
Synchronization
|
0° - 360°, 1° step or random
|
Trigger Mode
|
Automatic, manual or external trigger
|
CDN
|
Built-in single phase automatic CDN
Single-phase-3-line AC
220 V 16 A Max. 250 V
|
PFMF Tests as per IEC61000-4-8 with
optional module MFT 400/1200 or VMT series
|
Magnetic Field Strength
|
TCXS 111 Single-turn coil: 1 A/m -100 A/m (continuous)
100 A/m -400 A/m (1 s -10 s in short
time)
TCXS 113 Single-turn coil : 1 A/m -300 A/m (continuous)
300-1200 A/m (1 s -10 s in short time)
|
Current Distortion Rate
|
<5%
|
Current Waveform
|
50 Hz/60 Hz sine wave
|
Operating Current Range for Short Time
|
120 A-500 A
|
Continuous Operating Current Range
|
1 A-120 A
|
Test Duration
|
1 s -28800 s
|
Waveform Interval
|
1 s -9999 s
|
Coil Size
|
1000 mm *1000 mm or others
|
Trigger Mode
|
Auto, Manual, External signal trigger
|
Coil Turn
|
Single-turn /three-turn
|
Coil Shape
|
Rectangle or others
|
Output Magnetic Field Strength
|
Test Sequencing setup
|
Pulse Magnetic Field Tests as per IEC
61000-4-9 with optional module PMC 1200
|
Output Magnetic Field Range
|
100 A/m-1200 A/m
|
Waveform
|
Rise time: 6.4 μs(±30%); duration: 16 μs(±30%)
|
Output Current
|
100 A-2000 A
|
Polarity
|
Positive, negative, alternative
|
Synchronization
|
0°- 360°, 1° step or random
|
Pulse Cycle
|
5 s-99 s(the shortest time depends on test current)
|
Trigger Mode
|
Auto, manual, external signal trigger
|
Coil Shape
|
Rectangle
|
Number of Test
|
1-999
|
Model of Coil
|
TCXS 111
|
Coil Size
|
1000 mm *1000 mm
|
Output Magnetic Field Strength
|
Test sequencing setup
|
Coil Turn
|
Single
|
|
|
|
Power Fail Tests as per IEC 61000-4–11 (with
Optional Module VVT or VMT series) & IEC 61000-4-29
|
EUT Max. Voltage
|
250 V AC, 220 V DC
|
EUT Max. Current
|
AC/DC 16 A continuous current; AC 20 A continuous 5 s; 40 A continuous 3 s;
500 A impulse current
|
EUT Current and Voltage Measurement
|
LCD display, BNC terminal: voltage
100:1, current 10 A:1 V
|
Dip Level
|
0%-100%(for VVT/VMT series)
0%,40%,70%,80% (for
VVTxxxxSFseries)
|
Interruption Level
|
0%
|
Dip and Interruption Interval Time
|
5 ms-9999 ms
|
Dip and Interruption Duration
|
0.1-9999 cycle or 1 ms-9999 ms
|
Dip and Interruption Rise/Fall Time
|
1 µs -5 µs (100 Ω load)
|
Dip and Interruption Test Duration
|
1 s-9999 s
|
Time of Voltage Variation Decreasing
|
500 ms-9999 ms (50%-100%) or abrupt (same
with rise/ fall time for dips and
interruption)
1000 ms-9999 ms (0%-100%) or abrupt (same
with rise/ fall time for dips and
interruption)
|
Voltage Variation Level
|
0-100%
|
Duration after Voltage Variation
Decreasing
|
10 ms-99999 ms
|
Time of Voltage Variation Increasing
|
500 ms-9999 ms(50%-100%)
1000 ms-9999 ms(0%-100%)
|
Trigger Mode
|
Automatic, manual or external trigger
|
Synchronization
|
0°- 360°, 1° step or random
|
|
|
Models Selection
|
Name
|
Model No.
|
IEC 61000-4-4
|
IEC 61000-4-5
|
IEC 61000-4-8
|
IEC 61000-4-9
|
IEC 61000-4-11
|
IEC 61000-4-29
|
CDN single- phase-3-line
|
Compact
Immunity Test System
|
CCS
500
|
√
|
√
|
√
|
√
|
√
|
√
|
√
|
EFT/Burst CDN
|
EFTN 3816T
|
√
|
|
|
|
|
|
|
Surge CDN
|
SPN 3816T
|
|
√
|
|
|
|
|
EFT/Burst and Surge CDN
|
SEPN 3816T
|
√
|
√
|
|
|
|
|
SEPN 3832T
|
√
|
√
|
|
|
|
|
Power Frequency Magnetic Module
|
MFT 400
|
|
|
√
|
|
|
|
MFT 1200
|
|
|
√
|
|
|
|
AC Voltage Dips, Short Interruption and Variation
Module
|
VVT 2216S
|
|
|
|
|
√
|
|
VVT 2216SV
|
|
|
|
|
√
|
|
Power Fail and Power Frequency Magnetic Module
|
VMT 2216S
|
|
|
√
|
|
√
|
|
VMT 2216SV
|
|
|
√
|
|
√
|
|
Pulsed Magnetic Field Converter
|
PMC 1200
|
|
|
|
√
|
|
|