TOV test parameters |
||
Model |
TOV 380H |
TOV 230L |
Ucs |
AC 240V |
AC 240V |
Ut |
AC 1200V |
1.32 times of Ucs,1.55times of Ucs |
Ut current |
300A 200ms |
300A 5s |
Phase superposition |
0~360° |
-- |
Test supply |
Three phase,AC 380V 200kVA |
|
Main switch |
Silicon controlled |
|
Test box |
National standard size with SPD installation rail |
|
Material of test box |
Epoxy+ Organic glass sheet |
|
Measurement system |
PC remote control, current, voltage |
|
General parameters |
||
Working power supply |
Three phase,AC380V ± 10%,50/60Hz |
|
Environment temperature |
15~35℃ |
|
Relative humidity |
45%~75% |
|
Grounding resistance |
Less than 0.5Ω |
|
Dimensions |
1800×800×1800mm |
|
Weight |
Approx.700kg |
|
Waveform distortion rate <5% if without conductive dust, without fire, without explosion danger, without corroding metal and insulation gas for sine power supply waveform |