ESD-CALA is composed of calibration target BX602, coaxial line and adapter. BX602 is coaxial current target and designed as required by IEC/EN 61000-4-2 ed.2, which is used to monitor and judge whether the EDS pulse waveform output by ESD generators could meet the requirement of regulation or not. The additional attenuator (JZ602 or JZ603) is used for matching the output signal with the input capability of the oscilloscope
Features
> Meet the requirement of IEC61000-4-2 :2008
> 2 Ω impedance (± 5%)
> Test voltage up to 30 kV
> Insertion loss ± 0.5d B at 1 GHz
> Insertion loss ±1.2 dB at 4 GHz
Technical Parameters |
|
ESD test voltage |
Max. 30 kV |
Insertion loss |
±0.5 dB 1 GHz ±1.2 dB 4 GHz According to “target-attenuator-cable”, no need to test for target |
Installation method |
The target should be installed on the wall of shielded room or on a metal plate of 1.2m*1.2m at least |
Output terminal |
Coaxial SMA connector |
Attenuator |
According to the input power of oscilloscope, connect an attenuator externally to the output of BX602. |
Accessories |
|
ESD calibration target, coaxial line, adapter, user manual |
|
Options |
|
BX602-AD Calibration adapter |
50 ohm , used for BX602 verification;50 ohm (±2%),4 GHz, IEC/EN 61000-4-2 |
JZ602 attenuator |
20 dB |
JZ603 attenuator |
30 dB |