> 5.7 inch color touch screen front panel operation
> programmable test function
> Ethernet, RJ45 port for remote control, print and documentation
> imported electronic switch for main switch
General Parameters |
|
Display |
5.7-inch TFT touch screen |
Scope of Working Power Supply |
AC 110 V/220 V, ± 10%, 50 Hz/60 Hz (default AC 220 V in Chinese Mainland) |
Fuse |
6:00 AM |
Maximum Power Consumption |
100 W |
Communication Method |
Ethernet LAN |
User Storage Space |
Infinite (PC) |
Communication Method |
Ethernet, RJ 45 |
Auxiliary Interface |
D-sub 25 p |
External Trigger Input |
BNC, 5V TTL |
CROTriggers Output |
BNC, 5V TTL |
Operation Control Input |
BNC, 5V TTL |
Warning Light Output |
0 – 220 V AC 10 A/0 – 48 V DC 3A; The power supply depends on the type of external signal light |
Safety Circuit |
Stop working when the safety circuit is not closed |
Failure Detection |
When it fails, the front panel LCD displays and interrupts the instrument's operation |
Instrument Working Status Indication |
Front panel LED indication, LCD display |
Instrument Grounding Connection Method |
Use a flat grounding wire |
Chassis Size |
4U chassis (450 mm * 190 mm * 620 mm) |
Instrument Weight |
About 15 kg |
Ambient Temperature |
15℃~35℃ |
Relative Humidity |
45%~75% |
Atmospheric Pressure |
86 kPa~106 kPa |
Technical Parameters |
|
Compliant With IEC 61643-311 and GB/T 18802.311 Impulse Voltage Testing |
|
Test Voltage Range |
500 V - 5000 V ± 10% |
Voltage Waveform |
Wavefront time: 1000 V/μs ± 20% |
Source Output Impedance |
1000 Ω |
Output Polarity |
Positive and negative |
Pulse Cycle |
10 s – 99 s |
Test Frequency |
1-999 times |
Trigger Method |
Automatic, manual, external triggering |
Compliant With IEC 61643-311 and GB/T 18802.311 Impulse Voltage Testing |
|
Test Voltage Range |
500 V - 5000 V ± 10% |
Voltage Waveform |
Wavefront time: 1.2 μs ± 30% |
Source Output Impedance |
500 Ω |
Output Polarity |
Positive and negative |
Pulse Cycle |
10 s – 99 s |
Test Frequency |
1-999 times |
Trigger Method |