Technical Parameters |
|
Contact discharge |
200 V – 30,000 V (±5%) |
Air discharge |
200 V – 30,000 V (±5%) |
Voltage step |
100 V |
Hold time |
> 5 s |
Polarity |
Positive, negative |
RC module recognition |
Automatic recognition &LCD display |
Temperature and humidity |
Built-in temperature and humidity meter with functions of record and protection |
Repetition frequency |
Single /0.1/0.2/0.5/1/2/5/10/20 Hz |
Trigger mode |
Manual/automatic |
Pulse Counter |
1-9999 |
Contact discharge |
150pF/330 ohm |
Rise time |
0.8 ns +/- 25% |
Quick start test program |
Simple and easy to reset data online |
Evaluate procedure |
Contact discharge test at 20 Hz,evaluate the test points of DUT |
Standard Test Procedure |
accord with EN/IEC61000-4-2 |
General Parameters |
|
Weight |
Approx. 12.0 kg |
Weight (EDS gun) |
Approx. 2.0 kg |
Temperature |
15 ℃-35 ℃(operation condition) |
Humidity |
30%-60%(operation condition) |
Air pressure |
86 kPa-106 kPa |
Working power supply |
100 V - 240 V(±10%); 50/60 Hz(default AC 220 V 50 Hz for China Mainland) |
Standard Accessories |
||
1. Discharge Gun
|
Discharge gun and control 1 PCS |
|
2. High Voltage Generator
|
Programmable high voltage source 1 PCS |
|
3. Standard RC Modules EDS 20H-RC
|
150 pF / 330 ohm(default), removable |
|
4. Discharge Tip EDS 20H-CONT
|
Contact discharge tip |
|
5. Discharge Tip EDS 20H-AIR
|
Air discharge tip |
|
6. Grounding wire |
2 m |
|
7. Package |
Carrying case |
|
8. Documents |
Factory inspection report and user manual |
|
Optional Accessories |
||
1. RC Modules
|
150 pF/2000 ohm (EDS) 330 pF/2000 ohm (EDS) 330 pF/330 ohm (EDS) 100 pF/1500 ohm (EDS) 150 pF/150 ohm (EDS) 500 pF/500 ohm (EDS) |
500 pF/5000 ohm (EDS) 200 pF/0 ohm (EDS) 500 pF/0 ohm (EDS) 500 pF/100 ohm (EDS) Customization is available for RC module. |
2. 30 mm Air Discharge Tip EDS 30T-AIR (30mm)
|
For 15 kV to 30 kV test |
|
3. Battery Box EDS01
|
8*2600 mAh Ni-MH rechargeable batteries to replace AC source For contact discharge +30 kV, 1 Hz,approx. 8 h Dimension: 140 mm*100 mm*35 mm, weight 0.5 kg |
|
4. Battery Charger
|
Used for charging for battery box EDS 01 |
|
5. Electrostatic Discharge Desk
|
Grounded HV resistor 2 x 470 kohm |
|
Vertical coupling plane 500x500x3 mm |
||
Horizontal coupling plane 1600x800x2 mm |
||
Reference ground plane 2700x1800x2 mm |
||
Test table 1600x800x800 mm, solid wood,meet IEC 61000-4-2 requirements |
||
EDS CALA
|
Used for waveform calibration as per IEC/EN61000-4-2, including 2 ohm target /4 GHz,attenuator,coaxial cable,user manual |
|
7. Attenuator JZ602
|
Impedance: 50 ohm; Attenuation: 20 dB |
|
8. Attenuator JZ603
|
Impedance: 50 ohm; Attenuation: 20 dB |
|
9. EDS 20H-PC Communication EDS 20H-REMOTE |
Realize the remote control function of the upper computer software to the instrument |
|
10. PC Software Corelab |
> Compatible with Windows 7, Windows 8 and Windows 10 operating systems > support Ethernet interface and serial interface communication mode > support scheduling testing, test sequence parameters and the key test, the operation is simple > can be set up customer information, automatic generation of test reports, and support the export of test report, the convenient user record data in real time |