General Parameters |
||
Display |
5.7 inch TFT touch screen |
|
Power supply |
110 V / 220 V, ±10%, 50/60 Hz |
|
Fuse |
6 A |
|
Max. Power consumption |
100 W |
|
Memory space |
Infinite (PC) |
|
Communication mode |
Ethernet LAN、RJ45 |
|
External control |
D-sub 25p |
|
External trigger input |
BNC, 5 V TTL |
|
CRO output |
BNC, 5 V TTL |
|
Input |
BNC, 5 V TTL |
|
Synchronization |
45-65 Hz, 20-500 V between dotted terminal & undotted terminal |
|
Warning LED |
0-220 VAC 10 A/0-48 VDC 3 A Depending on external signal lamp |
|
Safety circuit |
Test only can be started when it is closed |
|
Failure test |
Display on LCD and interrupt the work at failure case |
|
EUT voltage monitoring output |
BNC,100:1 coaxial output |
|
EUT current monitoring output |
BNC,10A:1 V coaxial output |
|
Working indicator |
LED, LCD front panel display |
|
Grounding connection |
Flat ground line |
|
EUT power injection |
4 mm banana line |
|
Output mode |
Standard 1 m coaxial cable |
|
Size |
19 inch/6U/4U or 22U cabinet |
|
Weight(different weight for different model) |
Eg. PFS 2210SVD / PFS 2210SD / PFS 2216SD weight: 40 Kg; PFS 1130D weight: 24 Kg or 100 kg (with the power supply) |
|
Temperature |
15 ℃ ~ 35 ℃ |
|
Humidity |
45% ~ 75% |
|
Air pressure |
86 kPa ~106 kPa |
|
Power fail test as per IEC6100-4-11&IEC61000-4-29,GB/T17626.29 for single phase AC and DC including voltage dips, short interruption and voltage variation |
||
Power fail simulator |
||
EUT max. Voltage |
PFS22xxSD, PFS22xxSVD series: 250 V AC,300 V DC PFS05xxD series:55 V DC PFS11xxD series:130 V DC PFS12xxD series:140 V DC PFS22xxD series:300 V DC |
|
EUT max. current |
PFS2216SD、PFS2216SVD: 16 A AC/DC continuous current; 20 A continuous 5 s; 40 Acontinuous 3 s; PFS2210SVD、PFS2210SVD: 10 A AC/DC continuous current; 12 A continuous 5 s; 25 Acontinuous 3 s; PFSxx05D: 6 A PFSxx010D:12 A PFSxx20D: 22 A PFSxx25D: 28 A PFSxx30D: 33 A PFSxx50D: 55 A |
|
EUT power supply |
Single phase no more than 230 V 50~60 Hz±5% |
DC no more than 300 V DC Offer DC control |
EUT current and voltage measurement |
LCD display BNC terminal:voltage 100:1, current 100 A : 1 V |
|
Interrupt level |
0% |
|
Dips level |
0%-100%(for accessories VVT/VMT series) 0%,40%,70%,80%(for accessories VVTxxxxSF series ) |
|
Repetition rate for dips and interruption |
0.3-9999 cycle or 5 ms - 9999 ms |
|
Duration for dips and interruption |
50 ms – 50000 ms |
|
Test time for dips and interruption |
1 s – 9999 s |
|
Rise and fall time for dips and interruption |
Generator rated current ≤ 75 A: 1-5 μs (standard resistance calibration) Generator rated current >75 A: 1-50 μs (standard resistance calibration) |
|
Voltage variation level |
0-100% |
|
Time for decreasing voltage |
500 ms - 9999 ms(50%-100%)or abrupt(same with rise and fall time for dips and interruption) 1000 ms - 9999 ms(0%-100%)or abrupt(same with rise and fall time for dips and interruption) |
|
Time for increasing voltage |
500 ms - 60000 ms(50%-100%); 1000 ms – 60000 ms(0%-100%) |
|
Time for reduced voltage |
10 ms – 99999 ms |
|
DC voltage level |
0-120% |
|
DC voltage change continuous |
1 ms ~ 99999 ms |
|
DC voltage change interval |
50 ms ~ 50000 ms |
|
Synchronization |
0°~ 360°,1° step or random |
|
Trigger mode |
Automatic, manual or external |
|
Optional function |
Additional module for voltage variation test and two DC mains |
|
Peak impulse capacity of simulator |
||
Rated current |
Withstand current impulse capacity |
|
PFS 2210SD |
500 A |
|
PFS 2210SVD |
500 A |
|
PFS 2216SD |
500 A |
|
PFS 2216SVD |
500 A |
|
PFS 2232SD |
500 A |
|
PFS 2232SVD |
500 A |
|
PFS 0505D |
24 V/50 A, 48 V/100 A |
|
PFS 0510D |
24 V/50 A, 48 V/100 A |
|
PFS 1110D |
24 V/50 A, 48 V/100 A,110 V/220 A |
|
PFS 1120D |
24 V/50 A, 48 V/100 A,110 V/220 A |
|
PFS 1125D |
24 V/50 A, 48 V/100 A,110 V/220 A |
|
PFS 1130D |
24 V/50 A, 48 V/100 A,110 V/220 A |
|
PFS 1150D |
24 V/50 A, 48 V/100 A,110 V/220 A |
|
PFS 2210D |
24 V/50 A, 4 8V/100 A,110 V/220 A |
|
PFS 2225D |
24 V/50 A, 48V/100 A,110 V/220 A |
|
PFS 2250D |
24 V/50 A, 48 V/100 A,110 V/220 A |
|
Simulator with rated current no more than 16 A meets the load capacity specified in IEC 61000-4-11 |
||
Rated current |
Load current capacity |
|
PFS 2210SD PFS 2210SVD |
12 A AC/DC continuous current; 15 A continuous 5 s; 25 A continuous 3 s; |
|
PFS 2216SD PFS 2216SVD |
18 A AC/DC continuous current; 20 A continuous 5 s; 40 Acontinuous 3 s; |