The EFT 500 series is the multifunctional test simulator for EFT/Burst with a built-in single phase or three phase coupling /decoupling network. It is designed to meet the requirements of electronic and electrical standards. Stable performance and good reliability make the EFT 500 an optimum standalone solution for both testing during development and final testing of products and sub-assemblies. With internal automatic single phase or three-phase CDN, it is featured with good reliability, stable performance and easy use.
Features:
> 5.7 inch color touch screen front panel operation
> Internal automatic single phase or three-phase CDN
> Programmable setup of test parameters
> Controlled external three-phase high current CDN
> EFT 500S (single phase network) EFT 500T (three-phase network)
> Ethernet, RJ45 port for PC remote control, print and documentation
General Parameters |
|
Display screen |
5.7 inch TFT color touch screen |
Working mains |
AC 110 V/220 V,±10%,50/60 Hz,±5%(default AC 220V 50 Hz) |
Fuse |
6 A |
Power consumption |
200 W |
Memory space |
Infinite(PC) |
Communication mode |
Ethernet, RJ45 |
External trigger input |
BNC, 5 V TTL |
Operation control input |
BNC, 5 V TTL |
Failure detection |
LCD front panel display and Interrupt instrument work |
Working indicator |
LED, LCD front panel display |
Grounding mode |
Flat ground wire |
Size |
19 inch /4 U |
Weight |
Approx. 23 kg |
Temperature |
15 ~ 35 ℃ |
Humidity |
45% ~ 75% |
Air pressure range |
86kPa ~ 106kPa |
EFT/Burst test as per IEC 61000-4-4 |
|
Burst module |
|
Test voltage |
0.25 kV ~ 4.8 kV ±10% |
Polarity |
+/-/alternative |
Source impedance |
50 ohm |
Pulse waveform |
5/50 ns, 50 ohm load; 1000 ohm load |
Rise time |
5 ns±30%, 50 ohm load, 1000 ohm load |
Pulse duration |
50 ns±30%, 50 ohm load |
50 ns ~ 15/+100 ns, 1000 ohm load |
|
Pulse frequency |
0.1kHz ~ 1000 kHz |
Burst duration |
0.001ms ~ 1000ms |
Burst cycle |
11ms ~ 10100ms |
Test duration |
1 s ~ 9999 s |
Trigger mode |
Automatic, manual or external |
Synchronization |
0° ~ 360°, 1° step or random |
Network attenuation ratio |
<10% of setting voltage |
network |
EFT 500S built-in single CDN AC 250 V/16 A; DC-1 220 V/16 A; EFT 500T built-in three phase CDN AC 380 V/16 A; DC-1 220 V/16 A |
Standard Accessories |
Simulator, user manual, factory test report, test cable, power cord and two fuses |
Optional Accessories |
1. EFT CDN EFTN series IEC 61000-4-4 EFTN 3832T (three-phase AC 380V 32A) EFTN 38100T (three-phase AC 380V 100A) EFTN 6932T(three-phase AC 690V 32A) EFTN 69100T(three-phase AC 690V 100A) |
2. Capacitance coupling clamp CCC data signal line test |
3. Burst calibration devices, including TFB 500/TFB 1000, network port adapter and suit case: TFB 500: INPUT impedance 50 ohm, OUTPUT impedance 50 ohm, Attenuation ratio: 500:1 TFB 1000: INPUT impedance 1000 ohm, OUTPUT impedance 50 ohm, Attenuation ratio: 1000:1 |
4.PC control software Corelab With PC installed WIN 7, WIN 8 and WIN 10, it can be easily operated to make the measurement, based on the customized test program. It can identify any device with automatic configuration. It can easily generate test reports. |
Tel: +86 (0)512 6807 7192