The EVPG 20 is designed as per EHV-10 (load dump with rapid rate of change) and EHV-16 HV pulse specified in VW 80300:2016, which is used in vehicle HV electronic components test. It has 5.7-inch color touch screen for operation and also can be controlled by PC software.
Features:
> 5.7-inch color touch screen front panel operation;
> Output load dump with rapid rate of change and HV pulse;
> Built-in failure alarm function;
Technical Parameters |
|
Standard |
VW 80300-2016 (EHV-10-2) |
Waveform Feature |
Double Exponential Wave |
Pulse Voltage |
10 V ~ 50 V, ±20 % 50 V ~ 200 V, ±10 % |
Pulse Width |
≥50 μs(0 % ~ 0 %) |
Pulse Rise Time (0 % ~ 100 %) |
6.67 μs±20 % (3000 V/ms@20 V) |
Output Impedance |
2 Ω |
Pulse Interval |
1 ~ 99 s |
Pulse Polarity |
Positive only |
Counter |
1 ~ 99 |
Trigger Mode |
Manual/Auto/External Trigger |
Coupling Mode |
Capacitor |
Decoupling Mode |
Diode |
DUT Load |
DC 1000 V/100 A |
General Parameters |
|
Display Screen |
5/7 TFT touch screen |
Mains Supply |
AC 220 V,土10%,50/60 Hz |
Fuse |
6 A |
Max.Power Consumption |
200 W |
Dimension |
4U |
Weight |
Approx. 25 kg |
Relative Humidity |
15 ℃ ~ 35 ℃ |
Ambient Temperature |
45 % ~ 75 % RH (no condensing) |
Atmospheric Pressure |
86 kPa ~ 106 kPa |
Standard |
VW 80300-2016 (EHV-16) |
Waveform Feature |
Rectangular Wave |
Pulse Voltage |
10 V ~ 50 V, ±20 % 50 V ~ 200 V, ±10 % |
Pulse Width |
10 μs ~ 20 μs±20 % 20 μs ~ 200 μs±10 % |
Pulse Rise Time (0 % ~ 100 %) |
≤300 ns (<5 V/ns) |
Output Impedance |
2 Ω |
Pulse Interval |
0.1 ~ 10 s |
Pulse Polarity |
Positive only |
Counter |
1 ~ 999 |
Trigger Mode |
Manual/Auto/External Trigger |
Coupling Mode |
Capacitor |
Decoupling Mode |
Diode |
DUT Load |
DC 1000 V/100 A |
Accessories |
3-core power line, user manual, test line, flat grounded line, fuse*2 |
Optional Accessories |
Corelab test software Windows 7, Windows 8 and Windows 10 are supported, including various operating functions and standard libraries. Users can easily control the test equipment remotely, complete the test work, and automatically configure it to help users flexibly generate test reports. |