The CM 0301M is used for measuring alternating current and time domain pulse current in the frequency range from 10 Hz to 1 MHz. It features no waveform distortion and physically isolation with current carrier and transmits time harmonics current or pulse current to recording instruments in the form of voltage. It is capable of meeting the measurement requirement of time harmonics signal with frequency of 1 MHz or below and transient pulse with rise time less than 350 ns. Widely used in EMC conductive sensitivity (immunity) test.
s21 curve of CM 0301M
Technical Parameters |
|||
Sensitivity |
0.001 V/A |
High Frequency 3 dB cut-off |
1 MHz |
Output Resistance |
50 Ω |
Inner Diameter |
40 mm |
Max. Peak Current |
200,000 A |
Outer Diameter |
116 mm |
Max. rms Current |
400 A |
Thickness |
40 mm |
Rise Time |
350 ns |
Interface Type |
BNC type |
Current Time Product |
6 A·s |
Operating Temperature |
0 ℃ ~ 65 ℃ |
Low Frequency 3 dB cut-off |
10 Hz |
Weight |
1 kg |