Technical Parameters – EFT/Burst |
|
Single-phase Fully Automatic CDNs |
Comply with IEC/EN 61000-4-4 EFT/Burst immunity test. Maximum voltage up to 4.2 kV. Note: the actual output pulse voltage of CDNs is decided by the setting value of pulse generators. |
Phase Synchronization |
Any combination of L, N, PE with any phases and angles |
Coupling Route |
Any combination of L, N, PE |
Coupling Switching Mode |
Automatic switching, test sequencing |
Coupling Capacitor |
33 nF |
Coupling Attenuation |
<2 dB |
Residual Pulse Voltage of EUT Injection Port |
Not exceeding 10% of testing impulse voltage |
Technical Parameters - Surge |
|
Single-phase Fully Automatic CDNs |
Comply with IEC/ EN 61000-4-5 surge combination wave tests. Maximum voltage up to 6.0 kV (1.2/50 μs) and maximum current up to 3.0 kA (8/20 μs). Note: the actual output pulse voltage of CDNs is decided by the setting value of pulse generators. |
Phase Synchronization |
Any combination of L, N, PE with any phases and angles |
Coupling Route |
Any combination of L, N, PE |
Coupling Switching Mode |
Automatic switching, test sequencing |
Coupling Capacitor |
9 μF, 18 μF |
Coupling Resistor |
10 Ω, 0 Ω IEC standard mode or customized mode can be selected for coupling resistor. |
Coupling Attenuation |
<2 dB |
Residual Pulse Voltage of EUT Injection Port |
Not exceeding 15% of testing impulse voltage or two times of peak rated voltage of CDNs. |
General Parameters |
|
EUT Load Capacity |
Max. AC 450 V 16 A |
Max. DC 600 V 16 A |
|
EUT Power Switching Mode |
Automatic switching |
Working Power Source |
AC 110 V/220 V±10%,50 Hz /60 Hz, ±5% (AC 220 V 50 Hz in mainland China) |
Fuse |
6 A |
Max. Power Consumption |
200 W |
Auxiliary Interface |
D-sub 25p |
Working Status Indication |
LED indication on front panel |
Grounded Mode |
Flat grounded wire |
Dimension |
19"/6U |
Weight |
Approx. 50 kg |
Ambient Temperature |
15 ℃~35 ℃ |
Relative Humidity |
45% ~ 75% |
Atmospheric Pressure |
86 kPa ~ 106 kPa |
Accessories |
|
User Manual, Power Supply Line, Testing Line, Grounded Line, Fuse (spare part), Coaxial Line, CN 25 line |
|
Optional Generators & Calibration Tools |
|
EFT 500x |
EFT/Burst generator, as per IEC 61000-4-4, max. burst output voltage 4.8 kV |
EFT 600x |
EFT/Burst generator, as per IEC 61000-4-4, max. burst output voltage 6.0 kV |
EFT 500x |
EFT/Burst generator, as per IEC 61000-4-4, max. burst output voltage 7.0 kV |
CWS 600x |
Surge generator; as per IEC 61000-4-5; max. pulse output voltage 6.0 kV (1.2/50 μs), output current 3.0 kA (8/20 μs) |
CWS 800x |
Surge generator; as per IEC 61000-4-5; max. pulse output voltage 8.0 kV (1.2/50 μs), output current 4.0 kA (8/20 μs) |
CWS 1000x |
Surge generator; as per IEC 61000-4-5; max. pulse output voltage 10.0 kV (1.2/50 μs), output current 5.0 kA (8/20 μs) |
CCS 600x |
Surge and EFT/Burst generator; As per IEC 61000-4-4, max burst output voltage is 4.8 kV As per IEC 61000-4-5, max. pulse output voltage 6.0 kV (1.2/50 μs), output current 3.0 kA (8/20 μs) |
CCS 1000x |
Surge and EFT/Burst generator; As per IEC 61000-4-4, max burst output voltage is 4.8 kV As per IEC 61000-4-5, max. pulse output voltage 10.0 kV (1.2/50 μs), output current 5.0 kA (8/20 μs) |
VCF-80 |
HV differential probe, for calibration of Surge generator (open-circuit voltage waveform); test voltage max 8 kV, attenuation: 1000:1; |
TR 5025 |
HV current transducer, for calibration of Surge generator (short-circuit current waveform); Test current max 20 kA, attenuation 100:1; |
Calibration Kit for EFT/Burst Generators |
TFB 50: input impedance 50 Ω, output impedance 50 Ω, attenuation 55 dB; TFB 1000: input impedance 1000 Ω, output impedance 50 Ω, attenuation 60 dB; Supplied with network adaptors and tool box. |