Technical Parameters – EFT/Burst |
|
Three-phase Manual CDNs |
Comply with IEC/EN 61000-4-4 EFT/Burst immunity test. Maximum test voltage up to 4.2 kV. Note: the actual output pulse voltage of CDNs is subject to the setting value of generator. |
Coupling Route |
Manually switch, L1, L2, L3, N, PE random combination route |
Coupling Capacitor |
Manually switch, 33 nF |
Coupling Attenuation |
<2 dB |
Residual Pulse Voltage of EUT Injection Port |
Not exceeding 10% of testing impulse voltage |
Technical Parameters - Surge |
|
Three-phase Manual CDNs |
Comply with IEC/EN 61000-4-5 surge combination wave tests. Maximum voltage up to 6.0 kV (1.2/50 μs) and maximum current up to 3.0 kA (8/20 μs). Note: the actual output pulse voltage of CDNs is decided by the setting value of pulse generators. |
Coupling Route |
Manually switch, L1, L2, L3, N, PE arbitrary combination route |
Coupling Capacitor |
Manually switch, 9 μF, 18 μF |
Coupling Resistor |
Manually switch, 10 Ω, 0 Ω |
Coupling Attenuation |
<2 dB |
Residual Pulse Voltage of EUT Injection Port |
Not exceeding 15% of testing impulse voltage or two times of peak rated voltage of CDNs. |
General Parameters |
|
EUT Load Capacity |
Max AC 380 V 16 A, 50 Hz/ 60 Hz, 3-phase-5-line |
Working Power Source |
AC 110 V/220 V±10%,50 Hz /60 Hz±5% (AC 220 V 50 Hz in mainland China) |
Dimension |
19"/6U |
Weight |
Approx. 30 kg |
Ambient Temperature |
15 ℃ ~ 35 ℃ |
Relative Humidity |
45% ~ 75% |
Atmospheric Pressure |
86 kPa ~ 106 kPa |
Accessories |
||
1. User Manual |
2. Power Supply Line |
|
3. Testing Line |
4. Grounded Line |
|
5. Coaxial Line |
6. Short-circuit Plug |
|
Optional Generators & Calibration Tools |
||
EFT 500x |
EFT/Burst generator, as per IEC 61000-4-4, max. burst output voltage 4.8 kV |
|
EFT 600x |
EFT/Burst generator, as per IEC 61000-4-4, max. burst output voltage 6.0 kV |
|
EFT 500x |
EFT/Burst generator, as per IEC 61000-4-4, max. burst output voltage 7.0 kV |
|
CWS 600x |
Surge generator; as per IEC 61000-4-5; max. pulse output voltage 6.0 kV (1.2/50 μs), output current 3.0 kA (8/20 μs) |
|
CWS 800x |
Surge generator; as per IEC 61000-4-5; max. pulse output voltage 8.0 kV (1.2/50 μs), output current 4.0 kA (8/20 μs) |
|
CWS 1000x |
Surge generator; as per IEC 61000-4-5; max. pulse output voltage 10.0 kV (1.2/50 μs), output current 5.0 kA (8/20 μs) |
|
CCS 600x |
Surge and EFT/Burst generator; As per IEC 61000-4-4, max burst output voltage is 4.8 kV As per IEC 61000-4-5, max. pulse output voltage 6.0 kV (1.2/50 μs), output current 3.0 kA (8/20 μs) |
|
CCS 1000x |
Surge and EFT/Burst generator; As per IEC 61000-4-4, max burst output voltage is 4.8 kV As per IEC 61000-4-5, max. pulse output voltage 10.0 kV (1.2/50 μs), output current 5.0 kA (8/20 μs) |
|
VCF-80 |
HV differential probe, for calibration of Surge generator (open-circuit voltage waveform); test voltage max 8 kV, attenuation: 1000:1; |
|
TR 5025 |
HV current transducer, for calibration of Surge generator (short-circuit current waveform); Test current max 20 kA, attenuation 100:1; |
|
Calibration Kit for EFT/Burst Generators |
TFB 50: input impedance 50 Ω, output impedance 50 Ω, attenuation 55 dB; TFB 1000: input impedance 1000 Ω, output impedance 50 Ω, attenuation 60 dB; Supplied with network adaptors and tool box. |