PFS6950T3스마트형 삼상전압 하강 시뮬레이션
모델:

제품소개

전기망과 연결된 전기, 전자설비는 전기망 전압 하강, 숏타임 중단과 전압 변화의 영향을 받게 됩니다. 이와 같은 전압 변화가 나타나는 원인은 전기망, 변전설비 고장 또는 부하가 돌발적으로 큰 변동이 발생하기 때문입니다.
PDF가격문의

귀하 요구에 따른 온라인 맞춤제작>>

    Share
  • 제품기준
  • 제품특점
  • 제품파라미터
  • 제품부품
  • 판매후 서비스
  • 제품동영상
  • IEC/EN 61000-6-2
  • IEC/EN 61000-6-1
  • IEC/EN 61000-4-29
  • EN 61000-6-1
  • EN 61000-6-2
  • IEC 60601-1-2
  • IEC 61326
  • IEC/EN 61000-4-34


General Description

PFS 6950T3 simulates voltage dip, interruption and variation which are caused by power grid and transformer devices due to malfunction or sudden large change of load. It is required to test this kind of phenomena to check whether electronic and electrical device can sustain safety operating condition. PFS 2216 can conduct single-phase, two-phase or three-phase test. It is equipped with six controllable semi-conductor switches with protection function.

As Per:

IEC 61326

IEC 61000-4-29

IEC 61000-4-34

IEC 60601-1-2

EN 61000-4-34

EN 61000-6-1

EN 61000-6-2

Features:

Standalone test system

Full compliant three-phase test

Six switches with protection function

Rating voltage up to 3x690V

Max. Current for each phase up to 50A

Built-in installation for voltage increasing automatically

Operation via front panel

General Data

Display

5.7inch TFT touch screen

Power supply

110V/220V, ±10%, 50/60Hz

Max. power consumption

100W

External trigger input

5V TTL

CRO output

5V TTL

Input

5V TTL

Communication mode

Ethernet LAN, RJ45

Safety circuit

The test only can be started when it is closed

Size

19 inch/22 U or 35U

Weight

135kg

Temperature

15-35

Humidity

45% - 75%

Air pressure

86kpa- 106kpa

Technical Data

Model

PFS 6950T3

Output voltage range

0%-100%(three-phase or single-phase)

Max. input power supply

Three-phase AV 690V±5%, 50/60Hz ±5%

EUT max current

50A AC/DC continuous current;

500A impulse current

Interrupt level

0%

Dip level

0% - 100%

Repetition rate

0.1-9999 cycle or 1ms-9999ms

Duration

5ms – 9999ms

Test time

1s – 9999s

Rise and fall time

1 – 50us(100Ω load)

Interruption test

Test for single-phase, two-phase and three-phase

Dip test

L-N

L-L

IEC 61000-4-34: Figure3a, 3b,3c

Variation level

0% - 100%

Time for decreasing voltage

500ms – 9999ms (50%-100%) or 1-50us(100Ω load)

1000ms – 9999ms (0%- 100%) or 1-50us(100Ω load)

Time for increasing voltage

500ms – 9999ms (50% - 100%)

1000ms – 9999ms (0% - 100%)

Time for reduced voltage

10ms – 99999ms

Synchronization

0-360°, resolution 1°

Trigger

Auto, manual, external

EUT current and voltage measurement

LCD display

Calibration Parameters

Dip voltage

40%

70%

100%

Duration for dip

0.5 cycle

10 cycle

100 cycle

1000cycle

Phase angle for dip

0°

90°

180°

270°

360°

Rise/fall time

1us-50us with 100Ω load

Accessories

Three-core power cord

1

Four-core cable

2

Coaxial test cable

1

Calibration certificate

1

User manual

1

Test report

1